AK-100 Amorphous Silicon Reference Cell

Used to accurately calibrate solar simulator intensity for a-si type cells. Can be used individually or with the AK-110 µ-si reference cell for tandem type cells.
Request a Quote
AK-100 Amorphous Silicon Reference Cell Image by Konica Minolta Sensing Americas, Inc.

With the current increasing focus on renewable energy, solar cells (photovoltaic cells) are in the spotlight. As companies strive to improve the performance and characteristics of solar cells, it is necessary to perform evaluation of each product under standard test conditions. For accurate measurements, it is necessary to utilize a reference photovoltaic cell with a calibrated short-circuit current to adjust the intensity of the solar simulator to be used for measurements.

Konica Minolta Sensing’s filtered PV Reference Cells have the industries best accuracy and stability (in accordance with JIS C8904). The AK-100 can be used with the AK-110 for tandem type (a-si/µ-si) cells or as a standalone reference for a-si cells. All AK series Reference PV Cells come ready to use right out of the box with a PT-100 temperature sensor and a test report of short circuit current (Isc). They can also be delivered with calibration traceable to NREL.

Share
proddescription
Spectral mismatch error of less than 1%

By employing the high-accuracy optical filter technology cultivated for our color-measuring instruments, we have succeeded in greatly reducing spectral mismatch, which had been a major problem with conventional cells. Spectral mismatch can be suppressed to 1% or less, even when evaluating solar cell performance using JIS C-8942 Class C solar simulators.

Improved durability against solar simulator light

Standard reference cells using optical filters typically have low durability against exposure to solar simulator light.  This causes changes in characteristics due to the effects of light exposure (solarization) making accurate measurements difficult. AK series Reference PV Cells have a high durability against exposure to light, greatly reducing solarization and ensuring stability.

Improved reflection characteristics

Error in short-circuit current (Isc) due to multiple reflection is greatly reduced. Improvements in the optical structure suppress multiple reflections and reduce the 1.3% error of conventional products to 0.0%.

Maintain standard test condition of 25°C

A PT-100 temperature sensor is integrated into the package of the AK series cells, and can be connected to a temperature-controlled stage for reference PV cells to easily achieve and maintain the standard test condition (STC) of 25°C.

Calibration Certificate

A calibration certificate traceable to the National Renewable Energy Lab in the United States or Japan’s National Institute of Advanced Industrial Science and Technology is available as an optional accessory

Software for calculation of spectral mismatch

Every AK series cell includes our Spectral Mismatch Software.  The software can be used to determine the spectral mismatch between the AK Series PV Cell and any manufactured solar cell.  The software is available for download as a trial from our website.

Spectral mismatch calculation software download (free)

Share
Model AK-100
Spectral mismatch error * Within ±1%
Short-circuit current (Isc) at 25°±1°C 40 mA or higher
Receptor area 20 × 20 mm
Aperture area 47 × 47 mm
FOV 160°
FF 65% or higher
Temperature sensor PT-100
Connector for I-V measurement XLR-type Cannon plug (JIS standard)
Connector for temperature measurement LEMO connector (WPVS standard)
Operating temperature/ humidity range 20 to 30°C, relative humidity 65% or less
Storage temperature/ humidity range 15 to 30°C, relative humidity 30 to 75% with no condensation
Size (W × D × H) 120 × 90 × 21.5 mm
Weight 600 g
Short-circuit current Short-circuit current values measured under Konica Minolta standard test method are listed in the included test report
Standard accessories Storage Case PV-A01, M4 screws (4pcs), Connector for I-V Measurement (male) PV-A02, Protection Plate PV-A03, Test Report, Inspection Report
Optional Calibration Calibration Certificate (NREL, AIST Traceable)
* Spectral mismatch error evaluated according to Konica Minolta standard test conditions
 

 

Included Accessories

Spectral Mismatch Software

Spectral Mismatch Software

Email to a friend

Shop KMSA

Suggested Accessories

Reference Cell Temperature Controller

Reference Cell Temperature Controller

Email to a friend

Shop KMSA

AK-100 Amorphous Silicon Reference Cell Brochure: Download