CS-2000 Spectroradiometer
Konica Minolta $pageTitle

Click to Enlarge Thumbnails

Konica Minolta CS-2000 Spectroradiometer
Konica Minolta CS-2000 Spectroradiometer
Konica Minolta CS-2000 Spectroradiometer

The CS-2000 is a polychromatortype spectroradiometer which can measure contrasts of 100,000:1 and features the world's top level capability to detect extremely low luminance levels of 0.003 cd/m2.

The optical design and signal processing technologies of this instrument also provide accurate measurements of low to extremely low luminance and chromaticity measurements. Measurements are calculated in as little as 5 seconds, which eliminates mechanical and electrical noise factors.

The CS-2000 has a low polarization error of 2% which ensures more stable measurements of displays that use polarization, including LCDs. The selectable measuring angle feature enables the user to select the optimum angle of 1°, .2°, or .1° based on the application. With a temperature range from 5° to 35°C, this instrument is reliable in any work environment.

Measurements can be taken after a warm-up time of only 30 seconds. The CS-2000 has three measurement settings, internal and external synchronization, and multi-integration mode. Synchronization can be done either internally with user inputs or externally via line input. Multi-integration mode reduces variations due to unsynchronized measurement or synchronized measurements of sources that have irregular light emission cycles.

The CS-2000 features a color LCD screen and simple operation panel, easy for first time users.

PDFClick here for the CS-2000 Catalog.

PDFClick here for the CS-2000 Manual.


Features
  • Detects extremely low luminances - as low as 0.003 cd/m²
  • Easy operation with a color LCD screen
  • Measuring angle switching function
  • Low polarization error - Ideal for LCDs
  • Quick measurements
  • 3 measurement settings: internal synchronization, external synchronization and multi-integration mode

Specifications
Model CS-2000
Wavelength range 380 to 780 nm
Wavelength resolution 0.9 nm/pixel
Display wavelength bandwidth 1.0 nm
Wavelength precision +-0.3 nm (Median wavelength: 435.8 nm, 546.1 nm, 643.8 nm; Hg-Cd lamp)
Spectral bandwitdh 5 nm or less (half bandwidth)
Measuring angle (selectable)
1°                                         0.2°                                 0.1°
Measurement luminance range (Standard light source A)
0.003 to 5,000 cd/m² 0.075 to 125,000 cd/m² 0.3 to 500,000 cd/m²
Minimum measuring area
Φ5 mm ( Φ1 mm when using close-up lens) Φ1 mm ( Φ0.2 mm when using close-up lens) Φ0.5 mm ( Φ0.1 mm when using close-up lens)
Minimum measuring distance 350 mm (55 mm when using close-up lens)
Minimum spectral radiance display 1.0x10(-9) Ws/r, m(2), nm
Accuracy: Luminance *1(Standard light source A) +-2%
Accuracy: Chromaticity *1(Standard light source A)
x,y :+-0.003 (0.003 to 0.005 cd/m²)
x,y :+-0.002 (0.005 to 0.05 cd/m²)
x :+-0.0015 (0.05 cd/m² to )
y :+-0.001
x,y :+-0.003 (0.075 to 0.125 cd/m²)
x,y :+-0.002 (0.125 to 1.25 cd/m²)
x :+-0.0015 (1.25 cd/m² to )
y :+-0.001
x,y :+-0.003 (0.3 to 0.5 cd/m²)
x,y :+-0.002 (0.5 to 5 cd/m²)
x :+-0.0015 (5 cd/m² to )
y :+-0.001
Repeatability: Luminance(2σ)(Standard light source A) *2
0.4 % (0.003 to 0.05 cd/m²)
0.3 % (0.05 to 0.1 cd/m²)
0.15 % (0.1 to 5,000 cd/m²)
0.4 % (0.075 to 1.25 cd/m²)
0.3 % (1.25 to 2.5 cd/m²)
0.15 % (2.5 to 125,000 cd/m²)
0.4 % (0.3 to 5cd/m²)
0.3 % (5 to 10cd/m²)
0.15 % (10 to 500,000cd/m²)
Repeatability: Chromaticity(2σ)(Standard light source A) before *2
0.002 (0.003 to 0.005 cd/m²)
0.001 (0.005 to 0.1 cd/m²)
0.0006 (0.1 to 0.2 cd/m²)
0.0004 (0.2 to 5,000 cd/m²)
0.002 (0.075 to 0.125 cd/m²)
0.001 (0.125 to 2.5 cd/m²)
0.0006 (2.5 to 5 cd/m²)
0.0004 (5 to 125,000 cd/m²)
0.002 (0.3 to 0.5 cd/m²)
0.001 (0.5 to 10 cd/m²)
0.0006 (10 to 20 cd/m²)
0.0004 (20 to 500,000 cd/m²)
Polarization error 1°: 2 % or less(400 to 780 nm) 0.1° and 0.2°: 3 % or less(400 to 780 nm)
Integration time Fast: 0.005 to 16 sec.; Normal: 0.005 to 120 sec.
Measurement time 2 secs. minimum (Manual mode) to 243 sec. maximum (Normal mode)
Color space mode Lv x y, Lv u'; v';, Lv T Δuv, XYZ, spectral graph, dominant wavelength, excitation purity, scotopic luminosity (with CS-S10w Professional)
Interface USB 1.1
Operating temperature/ humidity range 5 to 35°C, relative humidity 80% or less with no condensation
Storage temperature /humidity range 0 to 35°C, relative humidity 80% or less with no condensation
Power AC adapter (100 to 240V~,50/60 Hz)
Current consumption Approx. 20 W
Size 158 (W) x 262 (H) x 392 (D) mm (Main Unite), ø70 x 95 mm (lens)
Weight 6.2 kg
*1 Average of 10 measurements in NOrmal mode at temperature of 23+- 2°C and a relative humidity of 65% or less
*2 10 measurements in NOrmal mode at a temperature of 23 +-2°C and a relative humidity of 65% or less

Included Accessories
AC Adapter AC-A312
AC Adapter (AC-A312)
Buy Now >
AC Adapter 120V
AC Adapter (120 V)
Buy Now >
CS-S10W Software Prof Vers. 1.6 USB
CS-S10W Software Prof Vers. (USB) 1.6
Buy Now >
AC Adapter Package
AC Adapter Package
Buy Now >
USB Cable - CS-A32
USB Cable - CS-A32
Buy Now >
AC Adapter 120V
AC Cable (120 V)

Suggested Accessories
Custom Illuminance Adapter for CS-2000
Custom Illuminance Adapter for CS-2000
Lens Cap CS-A31
Lens Cap CS-A31
Buy Now >
CS-A33 ND Filter
CS-A33 ND Filter (1/10)
Buy Now >
ND Filter 1/100 CS-A34
ND Filter (1/100) CS-A34
Buy Now >
CS-A35 Close-up Lens
CS-A35 Close-up Lens
Buy Now >

CS-A36 Adapter for CCD Camera
Buy Now >
CS-A33 ND Filter 1/10 w/ Calibration Certificate
CS-A33 ND Filter (1/10) w/Calibration Certificate
Buy Now >

Custom CS-2000 Focus Lock Ring
CS-A34 ND Filter 1/100 w/ Calibration Certificate
CS-A34 ND Filter (1/100) w/Calibration Certificate
Buy Now >
CS-A35 Close-up Lens w/ Calibration Certificate
CS-A35 Close-up Lens W/Calibration Certificate
Buy Now >
CS-A30 Storage Case
CS-A30 Storage Case
Buy Now >
Tripod & Pan Head Package
Tripod & Pan Head Package
Buy Now >
White Calibration Plate CS-A5 w/o Calibration Data
White Calibration Plate CS-A5 w/o Calibration Data
Buy Now >
White Calibration Plate CS-A5 with Calibration Data and Certificate
White Calibration Plate CS-A5 with Calibration Data and Certificate
Buy Now >
White Calibration Plate CS-A5 with Calibration Data
White Calibration Plate CS-A5 with Calibration Data
Buy Now >
CS-A1 ND Eyepiece Filter
CS-A1 ND Eyepiece Filter
Buy Now >

CaseCruzer (PELICAN) PN: CR2217-11
Quick Links

© 2006-2014 Konica Minolta Sensing Americas, Inc.

Designed and Maintained by Giovatto