The T-10 and T-10M illuminance meters are Lux meters used wherever the lux measurement, or illumination strength, of light striking a certain location, is needed for evaluation. The lux meter also has the ability to calculate the average values and illumination and display them on the LCD screen.

These lux meters enable comparison of illuminance values that can be displayed as a difference or percentage value. Target levels can be entered into the lux meter by measurement or manually using the keypad. The T-10 series of lux meters has the ability to automatically calibrate after switching on the device, measure continuous and intermittent light sources, and users can enter color correction factors for adaptation to certain light sources.

This lux meter features a very large measurement range of 0.01 to 299,900 lx and automatic range switching. The receptor head is removable to accommodate various types of measurements. Plus, users have the ability to connect multiple  up to 30  receptor heads through a serial port so measurements can be taken at several points simultaneously without moving the lux meter. This feature is especially convenient for large movie screens requiring the measurement of various areas simultaneously.

The T-10M lux meter has a 14mm diameter measurement area, ideal for tight spaces. The lux meter is battery operated for increased portability, and is capable of measuring flickering light.

PDFClick here for the T-10/T-10M Manual


Features
  • Easy to use  simple, multi-function operation, with keypad cover to avoid accidental operation
  • Automatic Calibration  automatically adjusts to zero at powering on for immediate measurement
  • Auto-range functions  enables the instrument to have the ability to measure a wide range, automatic or manual setting
  • LCD screen  backlit, LCD screen displays average values, illumination, comparison of illuminance values
  • Portable  small, compact, and powered by AA-size batteries
  • Versatile  also has the ability to measure flickering light

Specifications
Model

Illuminance Meter

T-10A

(Standard receptor head)

Illuminance Meter

T-10MA

(Mini receptor head)

Illuminance Meter

T-10W s A

(Waterproof mini receptor head)

Illuminance Meter

T-10W L A

(Waterproof mini receptor head)

Type

Multi-function digital illuminance meter with detachable receptor head (Multi-point measurements of 2 to 30 points is possible)

Illuminance meter class

Conforms to requirements for Class AA of JIS C 1609-1: 2006 Illuminance meters Part 1: General measuring instruments 

Conforms to DIN 5032 Part 7 Class B

Conforms to requirements for special Illuminance meters of JIS C 1609-1: 2006 *1

Receptor

Silicon photocell

Relative spectral response

Within 6% of the CIE spectral luminous efficiency V (<

Cosine response

Within 3%

Within 10%

Measuring range

Auto range (5 manual ranges at the time of analog output)

Measuring function

Illuminance (l x). Illuminance difference (l x). Illuminance ratio (%). Integrated illuminance ( l x h) time (h). average illuminance (l x).

Measuring range

Illuminance

0.01 to 299,900 l x; 0.001 to 29,990 fcd

1.00 to 299,900 l x; 0.1 to 29.990 fcd *2

Integrated illuminance

0.01 to 999,900 x 10 3 l x h 0.001 to 99,990 x 10 *3 fcd h / 0.001 to 9999 h

User calibration function

CCF (Color Correction Factor) setting function: Measurement value x 0.500 to 2.000

Linearity

+-2% +-1 digit of displayed value

Temperature/ humidity drift

Within +-3%

Digital output

USB

Analog output

1 mV/digit, 3 V at ma ximum reading; Output impedance: 10 K ; 90% response time: 28 ms

Display

3 or 4 Significant-digit LCD with backlight illumination (Automatic illumination)

Power source

2 AA-size batteries / ; AC adapter AC-A308 (optional; for 1 to 10 receptors) or AC adapter AC-A311 (optional; for 1 to 30 receptors)

Battery life

72 hours or longer (when alkaline batteries are used) in continuous measurement

Operating temperature/ humidity range

-10 to 40 deg C, relative humidity 85% or less (at 35 deg C) with no condensation

5 to 40 deg C, relative humidity of 85% or less (at 35 deg C) with no condensation

Storage temperature/ humidity range

-20 to 55 deg C, relative humidity 85% or less (at 35 deg C) with no condensation

0 to 55 deg C, relative humidity of 85% or less (at 35 deg C) with no condensation

Dimensions

69 x 174 x 35 mm

Main body: 69 x 161.5 x 30 mm

Receptor: O 16.5 x 13.8 mm

Cord length

-

1 m

5 m

10 m

Weight (without battery)

200 g (7.0 oz.)

205 g

260 g (Receptor head only: 120 g)

340 g (Receptor head only 200 g)

* 1 Conforms to requirements for Class AA of JIS C 1609-1:2006 for all items e xcept cosine response (f 2 ).

* 2 Although measurements below 1.00 l x are possible, they may not be stable due to the effects of electrical noise.

Notes regarding mini receptors and waterproof mini receptors;

* Do not touch the cable during measurements. Doing so may result in unstable measurement values.

* Secure the cable during measurements. Failure to do so may result in unstable measurement values.