CM-2500d Spectrophotometer
Click to Enlarge Thumbnails
The CM-2500d Spectrophotometer is a handheld, portable measurement instrument designed to evaluate the color and relative gloss characteristics of samples medium to large in size, particularly with flat surfaces. This high performance, precise spectrophotometer allows users to evaluate, reproduce, and control the color and appearance of objects in a more effective, streamlined process. Its advanced optical design and features help to establish and meet color quality standards more efficiently, maintain consistency in each batch of material, and communicate color seamlessly internally and throughout the supply chain. This makes the CM-2500d suited for each stage of the manufacturing process, including R&D analysis, quality control inspections, and process control procedures.
The CM-2500d Spectrophotometer is compatible with SpectraMagic NX software to record measurements and provide a more comprehensive color analysis.
Existing Customer?
If you're looking for service support or help with calibration then complete a service request form and we'll get right back to you
Features
- Fixed 8mm aperture is ideal for evaluating the color and appearance of samples medium to large
- Tight inter-instrument agreement measures consistently with other instruments of the same model, allowing color values and specifications to be communicated, shared, and coordinated seamlessly throughout the supply chain
- Pass/fail assessments immediately determines if the sample meets the defined standard or target color
- Numerical Gloss Control measures level of relative gloss for more control of a sample’s appearance
- Conveniently measures SCE and SCI simultaneously for more consistent readings and to accommodate the surface conditions of each sample
- Illuminated viewfinder conveniently illuminates where the instrument is positioned
- LCD displays measurement results clearly and reversing in direction according to how it’s held so that data is always facing upright to the user
- Its portable and lightweight powered by batteries or AC adapter
- Simple, straightforward navigation wheel with options easily understood by new users
- Six language options – English, Spanish, French, German, Italian, and Japanese
- integrates with SpectraMagic NX software to record measurements
Specifications
Model | CM-2500D |
---|---|
Model | CM-2500d |
Illumination / observation system | d/8 (diffuse illumination, 8° viewing ), equipped with simultaneous measurement of SCI (specular component included)/SCE (specular component excluded) Conforms to CIE No.15,ISO 7724/1,ASTM E1164,DIN 5033 Teil7 and JIS Z8722 Condition C standard. |
Sphere Size | Φ52mm |
Light-receiving element | Silicon photodiode array (dual 40 elements) |
Spectral separation device | Diffraction grating |
Wavelength range | 360nm to 740nm |
Wavelength pitch | 10nm |
Half bandwitdh | Approx. 10nm |
Reflectance range | 0 to 175%, resolution: 0.01% |
Light source | 2 pulsed xenon lamps |
Measurement time | Approx. 1.5 seconds |
Minimum interval between measurements | 3 seconds for SCI/SCE |
Battery performance | Alkaline manganese:approx. 1000 measurements |
Measurement/illumination area | MAV : Φ8mm/ Φ11mm |
Repeatability | Spectral Reflectance: Standard deviation within 0.1% (360 to 380nm within 0.2%) Colorimetric Value : Standard deviation within ΔE*ab 0.04(Measurement conditions:White calibration plate measured 30 x at 10-second intervals after white calibration was performed) |
Inter instrument agreement | within ΔE*ab 0.2 (MAV/SCI) Average for 12BCRA Series II color tiles compared to values measured with master body. |
UV adjustment | No UV adjustment |
Measurement mode | Single measurement/automatic averaging of multiple measurements(auto mode: 3, 5, 8 x/manual mode) |
Interface | RS-232C serial signals via RJ45 connector; USB 2.0 communication possible when using included USB-Serial Converter Cable IF-A24 |
Observer | 2/10 degrees (CIE 1931/2°,CIE 1964/10°) |
Illuminant | A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (simultaneous evaluation is possible using two light sources) |
Display data | Spectral value/graph, colorimetric value, color difference value/graph, PASS/FAIL result |
Color space / colorimetric data | L*a*b*, L*C*h, CMC (1:1), CMC (2:1), CIE94, Hunter Lab, Yxy, Munsell, XYZ, MI, WI (ASTM E313), YI (ASTM E313/ASTM D1925), ISO Brightness (ISO 2470), Density statusA /T, WI/Tint (CIE/Ganz), CIE00 |
Data memory | 1700 pieces of data (as SCI/SCE 1 data) * 700 pieces of data in the “defined in COND.” mode. * Total of the sample data for the COND and TASK modes and color diference target data |
Tolerance judgment | Tolerance for color difference (both box and eliptical tolerances can be set) |
Power source | 4 AA-size battery or AC adapter |
Size (W x;H x;D) | 69 x; 96 x; 193mm |
Weight | Approx. 670g (without batteries) |
Operating temperature / humidity range (*1) | 5 to 40°C, relative humidity 80% or less (at 35°C) with no condensation |
Storage temperature/humidity range | 0 to 45°C, relative humidity 80% or (at 35°C) with no condensation |
Standard accessories | White calibration plate, Target mask Φ8mm, Target mask Φ3mm , RS-232C cable, AC adapter,AA-size battery ( x;4) |
Optional Accessories | Hard case, Dust cover set, Dust cover, SpectraMagic trade;NX (software), Zero calibration box |
Suggested Accessories