ProMetric I Imaging Colorimeters


Click to Enlarge Thumbnails

The world’s fastest and most accurate high-resolution imaging colorimeters.

ProMetric® I imaging colorimeters are designed to address the demands for high-volume manufacturing of displays, backlit components, light sources, and device components. Applying tristimulus color filter technology based on standard CIE color-matching functions, ProMetric I provides highly accurate color and luminance measurements for lab testing and production-level inspection.

ProMetric I is designed around scientific-grade image sensors in a range of high-resolution options. These sensors enable pixel-level measurements of displays (LCD, OLED), inter- and intra-character luminance measurements on backlit keyboards and panels, and accurate measurements of LED luminance and color in luminaires or large LED arrays.

Existing Customer?

If you're looking for service support or help with calibration then complete a service request form and we'll get right back to you

Features

ProMetric I Family

The ProMetric® I family of imaging colorimeters offers the industry’s leading imaging solution for calibrated luminance and chromaticity measurement. The system’s internal tristimulus color filter provides an innate spectral response that aligns with standard CIE color-matching functions, while high-resolution imaging captures and analyzes millions of values at once across large fields of view and depths of field.

ProMetric I features Smart Technology™ features that simplify operation and provide more accurate measurement results. For example, the electronically controlled focus and aperture setting provides calibrated measurements over a wide range of working distances and field of view. This greatly simplifies measurement set up and allows fully-automated remote configuration of your production test systems

High-Resolution, Scientific-Grade Image Sensors

Each ProMetric I Imaging Colorimeter employs a scientific-grade image sensor optimized for low image noise to provide accurate, repeatable measurements. ProMetric cameras are designed to maximize resolution and dynamic range to ensure imaging performance. ProMetric I supports high-speed USB and Ethernet communications.

Available resolutions:

  • 2-megapixel (1600 x 1200)
  • 8-megapixel (3296 x 2472)
  • 29-megapixel (6576 x 4384)
  • 61-megapixel (9568 x 6380)

Purpose-Built for High-Speed Manufacturing

A production line is a harsh environment and reliable communications can be a challenge. ProMetric I supports high-speed USB and Ethernet communications, providing highly reliable operation over long distances, even in the most demanding manufacturing environments.

ProMetric I is optimized to work with the industry-leading TrueTest™ Automated Visual Inspection Software from Radiant. TrueTest provides a complete, turnkey solution for high-volume manufacturing of displays, illuminated keyboards and lighting products, with test sequencing interface to perform multiple inspections in-line in a matter of seconds. Whether you need to increase your test coverage or reduce test time for automated visual inspection, ProMetric I is designed specifically for your application.

Delivering Smart Technology™ Innovations

  • Smart Control™ for fast, precise setup: Smart Control allows you to electronically adjust both focus and aperture settings of your lens.
  • Smart Calibration™ for automatic high-accuracy results: ProMetric cameras offer a variety of electronically controlled lenses, each calibrated over a wide range of working distances and aperture settings. ProMetric systems monitor focal distance and aperture settings and automatically applies the correct flat field calibration, greatly simplifying setup and ensuring accurate measurement results.
  • Smart Touch™ to enable touch input via an integrated touchscreen display interface on the ProMetric I Imaging Colorimeter, which simplifies camera setup and control.

Calibrations

For highly accurate color and light measurements that match values perceived by the human eye, Radiant cameras and lenses are calibrated to obtain absolute data in any environment. Factory calibration procedures are performed on each Radiant camera to enable traceability to national standards such as NRL, ETL, PTB, and NIST.

Each Radiant camera and lens pair can also be calibrated for unique applications, in the factory or in the field, ensuring consistency regardless of variation in:

  • Focal Distance: Flat-field calibration; Image scaling
  • Device Spectral Power Distribution (SPD): Luminance calibration; Color calibrations, such as one-color calibration, four-color calibration (depending on RGB or single-color emitters); LED and white phosphor LED calibration
  • Lensing effects: Distortion calibration (if applicable)

ProMetric I Key Features and Highlights

  • High-speed, high-resolution image sensors, with better than 59dB dynamic range
  • CIE-matched color filters and neutral density filters, closely matches human visual perception
  • Multiple lens choices with Smart Calibration™ for a wide range of distance and aperture settings
  • Comes standard with Radiant ProMetric Software for light analysis and camera control
  • Works seamlessly with Radiant TrueTest Automated Visual Inspection test sequencing platform for automated applications

ProMetric I imaging colorimeter systems include Radiant Vision Systems ProMetric Software with API support, which provides complete measurement control and Point-of-Interest/Region-of-Interest functionality. In addition, the ProMetric I imaging colorimeters are backed by the industry’s best product warranty, including direct access to the Radiant Vision Systems technical support team to maximize the benefits for your application. For automated testing, ProMetric I is optimized to work with the industry-leading TrueTest Software — and specialized software modules—for high-volume manufacturing of displays, smartphones, notebooks and more.

Awards and Recognition

The ProMetric I Imaging Colorimeter has been honored with these awards for innovation:

Specifications

Parameter ProMetric I2 ProMetric I8 ProMetric I29 ProMetric I61
Primary Application

Uniformity Testing,
R&D Settings

Production, Display Testing, Pixel-level Measurement, Color Correction
Sensor Pixel Resolution 1600 x 1200 3296 x 2472 6576 x 4384 9568 x 6380
Sensor Megapixels 1.9 8.1 28.8 61.0
Sensor Type CCD CMOS
System Dynamic Range
(single exposure, per pixel)
59 dB (1 x 1 binning) 76 dB (1 x 1 binning)
Luminance (Minimum)* 0.00001 cd/mLimit of Detection
0.0001 cd/m@ SNR = 60
0.0005 cd/m@ SNR = 100
0.0005 cd/m2 Limit of Detection
0.0010 cd/m2 @ SNR = 60
0.0015 cd/m2 @ SNR = 100
Luminance (Maximum)

1010 cd/m2 with optional ND filters

System Accuracy** Illuminance ± 3%; Luminance (Y) ± 3%; Color Coordinates (x,y) ± 0.003
Short-term Repeatability* Illuminance ± 0.02%; Luminance (Y) ± 0.02%; Color Coordinates (x,y) ± 0.00005
Lens Type Electronically controlled focus and aperture
Focal Distances Available 24, 35, 50, 100, 200 mm 50, 100, 200 mm
Field of View
(Full Angle, H x V degrees)
24 mm 20° x 15°
35 mm 14° x 10°
50 mm 10° x 8°
100 mm macro 5° x 4°
200 mm 3° x 2°
24 mm 38° x 30°
35 mm 29° x 22°
50 mm 21° x 16°
100 mm macro 10° x 8°
200 mm 5° x 4°
50 mm 40° x 28°
100 mm macro 20° x 14°
200 mm 11° x 7°
Minimum Measurement Time*** 0.3 sec – photopic
1.1 sec – color
0.4 sec – photopic
1.2 sec – color
0.9 sec – photopic
2.4 sec – color
0.6 sec – photopic
1.8 sec – color
Spatial Measurement Capabilities Luminance, Radiance, Illuminance, Irradiance, Luminous Intensity, Radiant Intensity, CIE Chromaticity Coordinates, L*a*b* Color Scale, Correlated Color Temperature (CCT), Dominant Wavelength
Units

foot-lambert, cd/m2, nit, W/sr/m2, foot-candles, lux, lux-s, W/m2, W-s/m2, candela, W/sr. CIE (x, y) and (u’, v’), Kelvin (CCT)

Communication Interface Ethernet 100/1000, USB 2.0 10 Gigabit Ethernet (10 GigE)
Power 100-240 V, 50-60 Hz, 140 Watts
LCD Touch Panel Resolution: 800 x 600; Diagonal: 125 mm
Dimensions (H x W x D) 238 mm x 181 mm x 230 mm
Weight 4.9 kg 4.6 kg
Operating Temperature  0 – 30° C 5 – 35° C
Operating Humidity 20 – 70% non-condensing

*     Based on a virtual detector size of 100 x 100 pixels.

**   Based on illuminant A or user calibration for specific spectra. Based on a virtual detector size of 100 x 100 pixels and a minimum exposure time of 10ms.

*** For 100 cd/m2, using Ethernet.

Specifications subject to change without notice.

System Specifications:

  • Intel® Core™ i7-8086 CPU @ 4.00 GHz
  • 32 GB installed RAM

System Requirements:

  • Windows® 10, 64 bit
  • Ethernet 100/1000 or USB 2.0 port (I2, I8, and I29)
  • Desktop: PCI-E x8 lane slot (I61)
  • Laptop: Thunderbolt 3 Port (I61)