CS-200 Color and Luminance Meter
Click to Enlarge Thumbnails
Click Here to Watch Video
The CS-200 is designed to measure both luminance and chromaticity with high accuracy. It is a low cost option that is easy to use and features a new auto mode for determining measurement speed according the luminance of the subject being measured.
It can measure any light source over a wide range of luminance levels from 0.01 cd/m2 to 20,000,000 cd/m2. The selectable measuring angle feature enables the user to select the optimum angle of 1°, 0.2°, or 0.1° based on the application.
The CS-200 is lightweight and battery powered for optimum portability. Data management software CS-S10w is included as a standard accessory. The software enables the user to easily view and manage measurement data. Up to 101 measurement values can be stored in addition to 20 reference values.
Existing Customer?
If you're looking for service support or help with calibration then complete a service request form and we'll get right back to you
Features
- Portable, lightweight, compact and battery powered
- Easy to use and affordable
- Selectable measuring range
- Measure both luminance and chromaticity
- Auto mode to determine measurement speed
- Can be used for both low and high luminance
- Measuring angle switching
- CS-S10w data management software included as a standard accessory
- Eligible ISO 17025 Certification
Specifications
Model | CS-200 Luminance Color Meter |
---|---|
Measurement range | 0.01 – 200,000 cd/m² (Measuring angle 1 °) 0.01 – 5,000,000 cd/m² (Measuring angle 0.2 °) 0.01 – 20,000,000 cd/m² (Measuring angle 0.1 °) |
Accuracy (Measuring angle 1°) *1 (Standard Illuminant A; Temperature: 23 degC+-2 degC, Relative humidity: 65% max.) |
150 cd/m² ( for Illuminant A ) : Lv +-2 % +-1digit ; xy +-0.002 0.01-0.5 cd/m² ( for Illuminant A ) : Lv +-0.02cd/m² +-1digit ; 0.5-1 cd/m² ( for Illuminant A ) : Lv +-0.02 cd/m² +-1digit ; xy +-0.007 1-10 cd/m² ( for Illuminant A ) : Lv +-2 % +-1digit ; xy +-0.004 10-200,000 cd/m² ( for Illuminant A) : Lv +-2 % +-1digit ; xy +-0.003 5000 cd/m² ( for Illuminant A ) : color filter (R, G, B) ; xy +-0.006 Light source at 5000 cd/m² + color filter (R, G, B) |
Repeatability (Measuring angle 1°) *2 (Standard Illuminant A) | 0.01-1 cd/m² ( for Illuminant A ) : Lv +-0.01 cd/m² +-1digit ; — (2σ/Auto) 1-2 cd/m² ( for Illuminant A ) : Lv +-0.5 % +-1digit ; xy 0.002 (2σ/Auto) 2-4 cd/m² ( for Illuminant A ) : Lv +-0.5 % +-1digit ; xy 0.001 (2σ/Auto) xy +-0.007 4-8 cd/m² ( for Illuminant A ) : Lv +-0.5 % +-1digit ; xy 0.0005 (2σ/Auto) xy +-0.004 8-200,000 cd/m² ( for Illuminant A ) : Lv +-0.1% +-1digit ; xy 0.0004 ((2σ/Auto) xy +-0.003 |
Measurement time | Automatic setting between 1s and 60s (AUTO); Automatic setting to 1s or 3s (LTD. AUTO) 0.5 sec / meas. (Super-FAST) 1 sec / meas.(FAST)3 sec / meas. (SLOW) 12 sec / meas. (Super-SLOW) |
Measurement method | Spectral method, Grating + linear photo diode array |
Measuring angle | 1°, 0.2°, 0.1°(switchable) |
Minimum measuring area | 0.5 mm 0.1 mm (close up lens) |
Minimum measuring distance | 296 mm (Distance from front edge of metal lens barrel) |
Observer | 2/10° |
Color space | Lv x y, Lv u’ v’, Lv T Delta;uv, XYZ, dominant wavelength |
Measurement synchronization setting range | Vertical synchronization frequency : 40.00 to 200.00Hz |
Interface | USB 1.1 |
Power source | AC adapter or 4 AA-Size Batteries |
Battery life | Approx. 3 hours (continuous measurement / Fast mode / AA-size alkaline cells) |
Size | 95 mm (W) x 127 mm (H) x 330 mm (L) |
Weight | 1.8 kg (without battery) |
Operating temperature/humidity range | 0°C to 40°, relative humidity 85% or less (at 35°) with no condensation |
Storage temperature /humidity range | 0°Cto 45°C,relative humidity 85% or less (at 35°C)with no condensation |
*1 23°C+-2°C Lv = 0.01-10 cd/m², SLOW, average of 30 measurements Lv = 10 cd/m² and higher, SLOW, average of 10 measurements | |
*2 At 0.2°measuring angle, the amount of received light is approx. 1/25 of that for 1°. Therefore, the repeatability becomes the same as that for 1 degwith 25 times lower luminance. At 0.1°measuring angle, the amount of received light is approx. 1/100 of that for 1°. Therefore, the repeatability becomes the same as that for 1 degwith 100 times lower luminance. |